rambo domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /home/agsgnnfb/public_html/glorioustronics.com/wp-includes/functions.php on line 6131
rx-630SA determines total sulfur in petroleum products using the energy-dispersive X-ray fluorescence (EDXRF) method, which is quick, nondestructive, and economical. It does not need helium gas or a vacuum pump (Only an AC power cord is needed). The C/H correction calibration curve has been set at the factory, and the operator does not need to make a calibration curve for every product type.
Once a sample is set, the total sulfur is determined automatically in 300 sec (typical).
rx-630SA has a 12-position carrousel for higher productivity and a color touch screen for intuitive operation.
![]()
Specifications. CATALOG
| Test Methods | ASTM D4294-03, ISO 8754, JIS K 2541-4, etc |
|---|---|
| Type | 12 positions |
| Range | 15 mg / kg (LOD) to 6.00 wt % |
| Size ( W x D x H ) | 560 x 560 x 270 mm |
| Weight | 27 kg |
| Power Consumption | AC 100 to 240 V 200 VA (max) 50 / 60 Hz |