rambo domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /home/agsgnnfb/public_html/glorioustronics.com/wp-includes/functions.php on line 6131

acl-3 automatically determines 4 kinds of Colors – Saybolt, ASTM, Pt-Co/Hazen, and Gardner Colors, utilizing the Tri-Stimulus values from the spectroscopy method prescribed in ASTM D6045, which generates equivalent test results with conventional manual precision in about 5 seconds. Since no subjective judgment needs to be called, perfect traceability is achieved.
![]()
Specifications. CATALOG
| Test Methods | ASTM Color : ASTM D1500, D6045, ISO 2049
Saybolt Color : ASTM D156, D6045 Pt-Co / Hazen Color : ASTM D1209, D 5386, ISO 6271 Gardner Color: ASTM D1544, ISO 4630 |
|---|---|
| Measurement Method | Tri-stimulus values from the spectroscopy method |
| Photodetector | CMOS Linear image sensor |
| Measuring Wavelength | 400nm to 700nm, 10nm interval |
| Light Source Lamp | White LED (Lifetime: approx. 20,000h) |
| Display Panel | 5.7 inch TFT color LCD Touch Panel |
| Interface | USB port (for communication with PC)
USB memory (for file system: FAT32) Parallel port (for printer) |
| Size (W x D x H) | 540 x 383 x 255 mm |
| Weight | 15.5 kg |
| Power Consumption | AC 100 to 240 V 40 VA (max) 50 / 60Hz |